By visiting this site, you accept the use of cookies. More about our cookie policy.

GOST R 52951-2008

GOST R ISO 15353-2014 GOST P 55080-2012 GOST R ISO 16962-2012 GOST R ISO 10153-2011 GOST R ISO 10280-2010 GOST R ISO 4940-2010 GOST R ISO 4943-2010 GOST R ISO 14284-2009 GOST R ISO 9686-2009 GOST R ISO 13899-2-2009 GOST 18895-97 GOST 12361-2002 GOST 12359-99 GOST 12358-2002 GOST 12351-2003 GOST 12345-2001 GOST 12344-88 GOST 12350-78 GOST 12354-81 GOST 12346-78 GOST 12353-78 GOST 12348-78 GOST 12363-79 GOST 12360-82 GOST 17051-82 GOST 12349-83 GOST 12357-84 GOST 12365-84 GOST 12364-84 STATE STANDARD P 51576-2000 GOST 29117-91 GOST 12347-77 GOST 12355-78 GOST 12362-79 GOST 12352-81 GOST P 50424-92 STATE STANDARD P 51056-97 GOST P 51927-2002 GOST P 51928-2002 GOST 12356-81 GOST R ISO 13898-1-2006 GOST R ISO 13898-3-2007 GOST R ISO 13898-4-2007 GOST R ISO 13898-2-2006 STATE STANDARD P 52521-2006 GOST P 52519-2006 GOST R 52520-2006 GOST P 52518-2006 GOST 1429.14-2004 GOST 24903-81 GOST 22662-77 GOST 6012-2011 GOST 25283-93 GOST 18318-94 GOST 29006-91 GOST 16412.4-91 GOST 16412.7-91 GOST 25280-90 GOST 2171-90 GOST 23401-90 GOST 30642-99 GOST 25698-98 GOST 30550-98 GOST 18898-89 GOST 26849-86 GOST 26876-86 GOST 26239.5-84 GOST 26239.7-84 GOST 26239.3-84 GOST 25599.4-83 GOST 12226-80 GOST 23402-78 GOST 1429.9-77 GOST 1429.3-77 GOST 1429.5-77 GOST 19014.3-73 GOST 19014.1-73 GOST 17235-71 GOST 16412.5-91 GOST 29012-91 GOST 26528-98 GOST 18897-98 GOST 26529-85 GOST 26614-85 GOST 26239.2-84 GOST 26239.0-84 GOST 26239.8-84 GOST 25947-83 GOST 25599.3-83 GOST 22864-83 GOST 25599.1-83 GOST 25849-83 GOST 25281-82 GOST 22397-77 GOST 1429.11-77 GOST 1429.1-77 GOST 1429.13-77 GOST 1429.7-77 GOST 1429.0-77 GOST 20018-74 GOST 18317-94 STATE STANDARD P 52950-2008 GOST P 52951-2008 GOST 32597-2013 GOST P 56307-2014 GOST 33731-2016 GOST 3845-2017 STATE STANDARD P ISO 17640-2016 GOST 33368-2015 GOST 10692-2015 GOST P 55934-2013 GOST P 55435-2013 STATE STANDARD P 54907-2012 GOST 3845-75 GOST 11706-78 GOST 12501-67 GOST 8695-75 GOST 17410-78 GOST 19040-81 GOST 27450-87 GOST 28800-90 GOST 3728-78 GOST 30432-96 GOST 8694-75 GOST R ISO 10543-99 GOST R ISO 10124-99 GOST R ISO 10332-99 GOST 10692-80 GOST R ISO 17637-2014 GOST P 56143-2014 GOST R ISO 16918-1-2013 GOST R ISO 14250-2013 GOST P 55724-2013 GOST R ISO 22826-2012 GOST P 55143-2012 GOST P 55142-2012 GOST R ISO 17642-2-2012 GOST R ISO 17641-2-2012 GOST P 54566-2011 GOST 26877-2008 GOST R ISO 17641-1-2011 GOST R ISO 9016-2011 GOST R ISO 17642-1-2011 STATE STANDARD P 54790-2011 GOST P 54569-2011 GOST P 54570-2011 STATE STANDARD P 54153-2010 GOST R ISO 5178-2010 GOST R ISO 15792-2-2010 GOST R ISO 15792-3-2010 GOST P 53845-2010 GOST R ISO 4967-2009 GOST 6032-89 GOST 6032-2003 GOST 7566-94 GOST 27809-95 GOST 22974.9-96 GOST 22974.8-96 GOST 22974.7-96 GOST 22974.6-96 GOST 22974.5-96 GOST 22974.4-96 GOST 22974.3-96 GOST 22974.2-96 GOST 22974.1-96 GOST 22974.13-96 GOST 22974.12-96 GOST 22974.11-96 GOST 22974.10-96 GOST 22974.0-96 GOST 21639.9-93 GOST 21639.8-93 GOST 21639.7-93 GOST 21639.6-93 GOST 21639.5-93 GOST 21639.4-93 GOST 21639.3-93 GOST 21639.2-93 GOST 21639.0-93 GOST 12502-67 GOST 11878-66 GOST 1763-68 GOST 13585-68 GOST 16971-71 GOST 21639.10-76 GOST 2604.1-77 GOST 11930.7-79 GOST 23870-79 GOST 11930.12-79 GOST 24167-80 GOST 25536-82 GOST 22536.2-87 GOST 22536.11-87 GOST 22536.6-88 GOST 22536.10-88 GOST 17745-90 GOST 26877-91 GOST 8233-56 GOST 1778-70 GOST 10243-75 GOST 20487-75 GOST 12503-75 GOST 21548-76 GOST 21639.11-76 GOST 2604.8-77 GOST 23055-78 GOST 23046-78 GOST 11930.11-79 GOST 11930.1-79 GOST 11930.10-79 GOST 24715-81 GOST 5639-82 GOST 25225-82 GOST 2604.11-85 GOST 2604.4-87 GOST 22536.5-87 GOST 22536.7-88 GOST 6130-71 GOST 23240-78 GOST 3242-79 GOST 11930.3-79 GOST 11930.5-79 GOST 11930.9-79 GOST 11930.2-79 GOST 11930.0-79 GOST 23904-79 GOST 11930.6-79 GOST 7565-81 GOST 7122-81 GOST 2604.3-83 GOST 2604.5-84 GOST 26389-84 GOST 2604.7-84 GOST 28830-90 GOST 21639.1-90 GOST 5640-68 GOST 5657-69 GOST 20485-75 GOST 21549-76 GOST 21547-76 GOST 2604.6-77 GOST 22838-77 GOST 2604.10-77 GOST 11930.4-79 GOST 11930.8-79 GOST 2604.9-83 GOST 26388-84 GOST 14782-86 GOST 2604.2-86 GOST 21639.12-87 GOST 22536.8-87 GOST 22536.0-87 GOST 22536.3-88 GOST 22536.12-88 GOST 22536.9-88 GOST 22536.14-88 GOST 22536.4-88 GOST 22974.14-90 GOST 23338-91 GOST 2604.13-82 GOST 2604.14-82 GOST 22536.1-88 GOST 28277-89 GOST 16773-2003 GOST 7512-82 GOST 6996-66 GOST 12635-67 GOST 12637-67 GOST 12636-67 GOST 24648-90

GOST R 52951−2008 Palladium. Methods of atomic-emission analysis with arc excitation spectrum


GOST R 52951−2008

Group B59


NATIONAL STANDARD OF THE RUSSIAN FEDERATION

Palladium

METHODS OF ATOMIC-EMISSION ANALYSIS WITH ARC EXCITATION SPECTRUM

Palladium. Methods of arc atomic-emission analysis


OKS 39.060
AXTU 1709

Date of introduction 2009−07−01


Preface


The objectives and principles of standardization in the Russian Federation established by the Federal law of 27 December 2002 N 184-FZ «On technical regulation», and rules for the application of national standards of the Russian Federation — GOST R 1.0−2004 «Standardization in the Russian Federation. The main provisions"

Data on standard

1 DEVELOPED by the Open joint-stock company «Krasnoyarsk factory of nonferrous metals named after V. N. Gulidova» (OJSC «Krastsvetmet»), JSC «Ekaterinburg plant for processing non-ferrous metals"

2 SUBMITTED by the Technical Committee for standardization ТК102 «Platinum metals"

3 APPROVED AND put INTO EFFECT by the Federal Agency for technical regulation and Metrology dated 14 August 2008 N 175th St

4 INTRODUCED FOR THE FIRST TIME


Information about the changes to this standard is published in the annually issued reference index «National standards», and the text changes and amendments — in monthly indexes published information «National standards». In case of revision (replacement) or cancellation of this standard a notification will be published in a monthly information index «National standards». Relevant information, notification and lyrics are also posted in the information system of General use — on the official website of the Federal Agency for technical regulation and Metrology on the Internet

1 Scope


This standard covers refined palladium bullion and powder with a mass fraction of palladium is not less than 99.8%, designed for the production of alloys, semi-finished products, chemical compounds of palladium.

Chemical composition of refined palladium bullion and powder meets the requirements of GOST R 52244.

The standard sets the spectrographic method atomic emission analysis (with an arc excitation spectrum) for determining the mass fraction of impurities: aluminum, iron, gold, iridium, calcium, cobalt, silicon, magnesium, manganese, copper, Nickel, tin, osmium, platinum, rhodium, ruthenium, lead, silver, antimony, tellurium, chromium, zinc and spectrometric method of atomic emission analysis (with an arc excitation spectrum) for determining the mass fraction of impurities, aluminum, bismuth, iron, gold, iridium, calcium, cobalt, silicon, magnesium, manganese, copper, Nickel, tin, osmium, platinum, rhenium, rhodium, ruthenium, lead, silver, antimony, tellurium, titanium, chromium and zinc in affilirovannomu palladium.

2 the Essence of the methods


Methods of analysis based on the evaporation and excitation of the atoms of the sample in arc discharge, measuring the intensity of radiation of atoms determined of impurity elements and subsequent determination of the mass fraction of these elements using the calibration dependences obtained by standard samples of composition of palladium.

3 Normative references


This standard uses the regulatory references to the following standards:

GOST R 8.563−96 State system for ensuring the uniformity of measurements. Methods of measurement

GOST R ISO 5725−1-2002 Accuracy (trueness and precision) of methods and measurement results. Part 1. General provisions and definitions

GOST R ISO 5725−3-2002 Accuracy (trueness and precision) of methods and measurement results. Part 3. Intermediate indicators the precision of a standard measurement method

GOST R ISO 5725−4-2002 Accuracy (trueness and precision) of methods and measurement results. Part 4. The main methods of determining the correctness of a standard measurement method

GOST R ISO 5725−6-2002 Accuracy (correctness and precision) of methods and measurement results. Part 6. The use of precision values in practice

GOST R 52244−2004 Palladium refined. Specifications

GOST R 52599−2006 Precious metals and their alloys. General requirements for methods of analysis

GOST 6709−72 distilled Water. Specifications

GOST 14261−77 hydrochloric Acid of high purity. Specifications

GOST 18300−87 ethyl rectified technical. Specifications

GOST 24104−2001 laboratory Scales. General technical requirements

GOST 25336−82 Glassware and equipment laboratory glass. The types, basic parameters and dimensions

Note — When using this standard appropriate to test the effect of reference standards in the information system of General use — on the official website of the Federal Agency for technical regulation and Metrology on the Internet or published annually by the information sign «National standards» published as on January 1 of the current year and related information published monthly indexes published in the current year. If the reference standard is replaced (changed), when using this standard should be guided by replacing (amended) standard. If the reference standard is cancelled without replacement, then the situation in which the given link applies to the extent that does not affect this link.

4 Terms and definitions


This standard applies the terminology according to GOST R ISO 5725−1 and GOST R 8.563.

5 Spectrographic method of atomic-emission analysis with arc excitation spectrum


When the spectrographic method, using photographic registration of the emission spectra.

This method allows the determination of mass fraction of impurity elements in the ranges given in table 1.


Table 1 — Ranges of mass fraction of detectable elements

Percentage

     
The name of the element Measurement range of mass fraction of
  DC
AC
Aluminium From 0.0005 to 0,0040 incl.
From 0,0010 to 0.020 incl.
Iron From 0.0005 to 0,0040 incl.
From 0.0020 to 0.050 incl.
Gold From 0.0005 to 0,0040 incl.
From 0,0010 to 0.050 incl.
Iridium From 0,0010 to 0.020 incl.
From 0.0020 to 0.10 incl.
Calcium From 0.0005 to 0,010 incl.
From 0,0010 to 0.020 incl.
Cobalt From 0,0010 to 0.020 incl.
From 0,0010 to 0.020 incl.
Silicon From 0.0005 to 0,0040 incl.
From 0,0010 to 0.020 incl.
Magnesium From 0.0005 to 0,0040 incl.
From 0,0010 to 0,010 incl.
Manganese From 0.0005 to 0,010 incl.
From 0,0010 to 0.020 incl.
Copper To from 0,0005 0,0010 incl.
From 0,0010 to 0,010 incl.
Nickel From 0.0005 to 0,0040 incl.
From 0,0010 to 0,010 incl.
Tin From 0.0005 to 0,0040 incl.
From 0,0010 to 0.020 incl.
Osmium From 0,0010 to 0,010 incl.
From 0.0020 to 0.010 is incl.
Platinum From 0,0010 to 0.020 incl.
From 0.0020 to 0.10 incl.
Rhodium From 0,0010 to 0.020 incl.
From 0.0020 to 0.10 incl.
Ruthenium From 0,0010 0,0040 to incl.
From 0.0020 to 0.10 incl.
Lead From 0.0005 to 0.020 incl.
From 0,0010 to 0.020 incl.
Silver To from 0,0005 0,0010 incl.
From 0.0005 to 0,010 incl.
Antimony From 0.0005 to 0.020 incl.
From 0,0010 to 0.020 incl.
Tellurium From 0.0005 to 0.020 incl.
From 0,0010 to 0.020 incl.
Chrome From 0,0010 to 0.020 incl.
From 0,0010 to 0.020 incl.
Zinc From 0.0005 to 0,010 incl.
From 0,0010 to 0.020 incl.

5.1 Accuracy (trueness and precision)

5.1.1 indicators of the accuracy of the method

Indicators of accuracy of the method: limit of the interval in which with a probability ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраof 0.95 is the absolute error of the results of the analysis (assigned error) ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, standard deviation of repeatability ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, the values of the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand reproducibility limit ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраdepending on the mass fraction of the element-impurities are shown in table 2.


Table 2 — indicators of the accuracy of the method (ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра0,95)

Percentage

             
Level mass fraction
the definition
given elements

The boundaries of the interval of absolute error
±ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The standard deviation of repeatability ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The standard deviation of intermediate precision
ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The limit of intermediate precision

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

Limit
of rehabilitation
of conductivity ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,0005 0,0004
0,0002 0,0007 0,0002 About 0.0006 0,0007
0,0008 About 0.0006
0,0003 0,0010 0,0003 0,0008 0,0010
0,0010 0,0011
0,0004 0,0014 0,0005 0,0015 0,0018
0,0020 0,0029
0,0008 0,0030 0,0015 0,0042 0,0050
0,0030 0,0035
0,0011 0,0040 0,0018 0,0050 0,0060
0,0050 0,0047
0,0017 0,0060 0,0024 0,0067 0,0080
0,010 0,007
0,002 0,007 0,004 0,010 0,012
0,020 0,011
0,003 0,010 0,005 0,015 0,018
0,030 0,013
0,003 0,012 0,007 0,019 0,023
0,050 0,017
0,004 0,015 0,008 0,023 0,028
0,08 0,02
0,007 0,02 0,011 0,03 0,04
0,10 0,02
0,008 0,03 0,012 0,03 0,04



For intermediate values of the mass fraction values in terms of accuracy find using linear interpolation according to the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, (1)


where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis the result of the analysis;

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраthe lower and upper bounds of a subrange of the mass fraction, which is the result of the analysis;

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — the value of the indicator of accuracy, corresponding to the lower and upper bounds of the subrange of the mass fraction, which is the result of the analysis;

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — the value of the index of precision for test results ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

.

5.1.2 Correctness

To estimate the systematic error of this method for the determination of impurity elements in palladium are used as reference values certified values of mass fractions of elements in the state standard samples of composition of GSO 7615−99 of palladium (set PD-36), GEO 7331−96 (set PD-28) or other GEO, the next set of detectable elements and metrological characteristics.

Systematic error of the method at the significance level ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра5% nonsignificant according to GOST R ISO 5725−4 for all defined elements-admixtures in palladium at all levels of the designated contents.

5.1.3 Precision

5.1.3.1 the Range of results four definitions obtained for the same sample by one operator using the same equipment within the shortest possible time intervals that may exceed the specified in table 2, the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраfor ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра4 according to GOST R ISO 5725−6 on average not more than once in 20 cases with proper use of the method.

5.1.3.2 within a single laboratory result analysis of the same sample obtained by different operators using the same equipment on different days, can vary with the excess specified in table 2 of the limit of intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраaccording to GOST R ISO 5725−3 on average not more than once in 20 cases with proper use of the method.

5.1.3.3 results of the analysis of the same samples obtained by two laboratories may vary with the excess indicated in the table 2 reproducibility limit ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраaccording to GOST R ISO 5725−1 on average not more than once in 20 cases with proper use of the method.

5.2 Requirements

5.2.1 General requirements and safety requirements

General requirements for method of analysis, the requirements for security of works and environmental safety — according to GOST R 52599.

5.2.2 qualifications of performers

To perform analysis allowed persons over the age of 18, trained in the prescribed manner and allowed to work independently on used equipment.

5.3 measurement Means, auxiliary devices, materials and reagents


Spectrograph with diffraction being a system of condensers, designed to produce spectra in the range of 240−350 nm, reciprocal linear dispersion of 0.6−0.7 nm/mm.

Arc generator of AC or DC power up to 15 A.

Microdensitometer designed to measure the optical density (blackening) of spectral lines.

Laboratory scales according to GOST 24104 with a limit of permissible absolute error of less than ±0.005 g.

Electrodes of graphite [1] 6 mm in diameter with a crater depth of 1−3 mm and a diameter of 4 mm.

Electrodes of graphite [1] 6 mm in diameter, sharpened to a hemisphere or truncated cone.

Photographic plates of the spectral for normal blackening in the spectral range 240−350 nm.

Contrast developer and fixer for photographic plates.

Electric stove with a closed spiral.

Glasses chemical resistant according to GOST 25336.

Hydrochloric acid O. S. h according to GOST 14261.

Rectified ethyl alcohol according to GOST 18300.

Distilled water according to GOST 6709.

Samples for calibration (samples of palladium with the previously established values of the mass fraction of impurity elements).

Standard samples of composition of palladium with an error of certified values mass fraction of impurities not exceeding 1/3 of the value of the absolute error of the present method for each level in mass fraction.

Allowed the use of other measurement means, auxiliary devices, materials and reagents, subject to obtaining accuracy rate that rivals those in table 2.

5.4 Sampling and sample preparation

5.4.1 the Selection of laboratory samples from ingots or powder of refined palladium is carried out according to GOST R 52244.

5.4.2 Samples of palladium can enter the analysis in the form of ribbon, wire, shavings, sponge, powder.

5.4.3 Samples received for analysis in the form of tape, wire or shavings for the removal of surface contamination are placed in a glass beaker and is boiled in hydrochloric acid solution, diluted 1:1 for 10−15 min. the resulting solution was decanted, the sample was washed with distilled water by decantation 4−5 times and dried in air. Samples of the powder and sponge not treated with acid.

5.4.4 From the palladium sample received for analysis selected 4 sample, from the samples to the calibration or standard samples — 2 samples weighing between 100 to 150 mg each, depending on depth electrodes used, but are the same for one spectrogram. The sample in powder form are pressed into the crater of the graphite electrode.

5.5 Preparation of equipment for measurement

5.5.1 Equipment ready to work according to the instructions. The wavelengths of the analytical lines and lines of comparison, the working modes of the instruments recommended to perform the analysis presented in tables 3 and 4, respectively. For each designated item, choose one of the recommended wavelengths. The use of other lines and operating modes subject to receipt of the accuracy rate that rivals those in table 2.


Table 3 — wavelengths of the analytical lines

In nanometers

     
The name of the element Wavelength of analytical line
Wavelength comparison lines (palladium)
Aluminium 309,27
Background, 306,61
Iron 302,06
Background, 302,17
Gold 267,59
Background, 268,62
Iridium 266,47
Background, 268,62
  322,0
Background, 321,89
Calcium 315,89
Background, 321,89
Cobalt 241,16
Background, 268,62
  345,35
Background, 302,17
Silicon 288,15
Background, 302,17
Magnesium 285,21
Background, 302,17
  280,2
Background, 302,17
Manganese 259,37
Background, 268,62
  260,57
Background, 268,62
Copper 324,75
Background, 321,89
  327,3
Background, 321,89
Nickel 341,47
Background, 332,09
  Of 301.2
Background, 302,17
Tin 286,33
Background, 302,17
  To 317.5
Background, 321,89
Osmium 283,86
Background, 268,62
Platinum 270,58
Background, 268,62
  299,4
Background, 302,17
Rhodium 332,30
Background, 332,09
Ruthenium 366,13
Background, 356,66
  298,89
Background, 302,17
Lead 283,30
Background, 302,17
Silver 328,06
Background, 328,72
  338,2
Background, 332,09
Antimony 287,7
Background, 302,17
Tellurium 238,5
Background, 268,62
Chrome 284,32
Background, 268,62
  286,51
Background, 268,62
Zinc 334,5
Background, 332,09
Note — the Background means that the comparison lines chosen wavelength corresponding to the minimum of the blackening of photographic plates near the analytical line of the designated element (either side, but with the same for a given element in all the spectra on the photographic plate).



Table 4 — Recommended operation modes

   
The name of the parameter The value of the parameter
DC-arc:
 
the strength of the current, And
10−12
Alternating current arc:
 
the frequency of discharge in Hz
100
a fixed value of the phase of ignition
60°
the strength of the current, And
10−12
Conditions of photographing of the spectra:
 
gap width, mm
0,015
the exhibition, with
60

5.5.2 electrode holders and devices are cleaned with alcohol from surface contamination.

5.5.3 Include water cooling of the electrode holders.

5.5.4 Prepared for analysis a sample of palladium was placed in a graphite crater electrode. Contradictator is a graphite rod sharpened to a hemisphere or truncated cone. When used as a driving source of a DC arc of the analyzed sample is the anode.

5.5.5 the Interelectrode gap set on the magnified image of the arc on the screen, the intermediate diaphragm of 5 mm and maintain strictly constant, adjusting it throughout the exposure.

5.6 measurements

5.6.1 To obtain a calibration curve using standard samples of composition of palladium or samples for calibration. Spectra of each standard sample (sample for calibration) and a test sample is photographed under the same conditions. For each standard sample (for calibration) get two, and for the sample analyzed four spectrograms. When the mass fraction of the elements in the sample exceeding the upper limit of the interval content using a constant current, specified in table 1, the photographing of the spectra is carried out again by use of AC.

5.6.2 photographic Plates showing, rinsed in water, fixed, washed in running water and dried.

5.6.3 using microdensitometry on each spectrogram to measure the blackening of analytical lines of the designated element ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand the surrounding background ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand calculate the difference of pochernenija ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра. The values to ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраpass to values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраfrom table A. 1 given in Appendix A. Using the values of ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраobtained for standard samples to build a calibration curve in the coordinates: ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраwhere ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — mass fraction of the element in the standard sample (the sample for calibration),

%.

5.6.4 In the upper end of the range of mass fractions allowed the construction of calibration graphs in the coordinate: ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis the difference of pochernenija analytical lines and comparison lines (palladium).

5.6.5 According to the calibration schedule, with four parallel values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраor ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, respectively, obtained four spectrograms for each sample, find the four results of parallel measurements of the mass fraction of each element-impurities in the sample.

5.7 evaluation of the acceptability of the results of parallel measurements and obtaining the final result of the analysis

5.7.1 acceptance of the results of parallel measurements was evaluated in accordance with GOST R ISO 5725−6 by mapping the range of these results ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраwith the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраshown in table 2.

5.7.2 If the range of the results of four parallel measurements ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраdoes not exceed the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, all the results and recognise acceptable for the final result of the analysis take the average of results of four parallel measurements.

5.7.3 If the range of the results of four parallel measurements exceeds ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, spend four parallel determinations.

Expect the critical range for up to eight parallel measurements ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраaccording to the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, (2)


where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis the standard deviation of repeatability.

If the received eight parallel measurements the value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраdoes not exceed the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, as a final result of the analysis taking the arithmetic mean of results of eight parallel measurements. Otherwise, as a final result of the analysis taking the median of the results of eight parallel measurements, if the regulations of the enterprise stipulate otherwise.

5.8 the Control of accuracy of analysis results

5.8.1 Control intermediate precision and reproducibility

When monitoring intermediate precision (variable factors of the operator and time) is the absolute difference of the two results of the analysis of the same sample obtained by different operators using the same equipment on different days, should not exceed the limit of intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраspecified in table 2.

Under the control of the reproducibility of the absolute disparity of the two results of the analysis of the same samples, obtained in two laboratories, in accordance with the requirements of this standard must not exceed the limit of reproducibility ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраshown in table 2.

5.8.2 verification of correct

The checking is carried out by analyzing standard samples of composition of palladium. The samples used for control of correctness cannot be used to obtain the calibration dependencies.

When checking the difference between the analysis result and the accepted reference (certified) value of the mass fraction of element-impurities in the standard sample should not exceed a critical value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

The critical value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis calculated by the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, (3)


where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — the error of setting the reference (certified) value of mass fraction of element-impurities in the standard sample;

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — the boundaries of the interval of the absolute error of the result of the analysis (values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраshown in table 2).

6 Spectrometric method of atomic-emission analysis with arc excitation spectrum


When the spectrometric method, using a photoelectric method of recording emission spectra.

The method allows to determine the mass fraction of impurity elements in the ranges given in table 5.


Table 5 — Ranges of mass fractions of the identified elements

Percentage

   
The name of the element Measurement range of mass fraction of
Aluminium From from 0.0002 to 0.040 incl.
Bismuth From 0,00010 to 0,010 incl.
Iron From 0,00010 to 0.10 incl.
Gold From from 0.0002 to 0.050 incl.
Iridium From 0,0010 to 0.10 incl.
Calcium From 0.0004 inch to 0.020 incl.
Cobalt From 0,00010 to 0,010 incl.
Silicon From from 0.0002 to 0.040 incl.
Magnesium From 0,00010 to 0,030 incl.
Manganese From 0,00010 to 0,010 incl.
Copper From 0,00010 to 0,030 incl.
Nickel From 0,00010 to 0.050 incl.
Tin From 0,00010 to 0.020 incl.
Osmium From 0,0003 0,010 to incl.
Platinum From 0,0010 to 0.10 incl.
Rhenium From 0.0005 to 0,0060 incl.
Rhodium From 0,0010 to 0.10 incl.
Ruthenium From 0,0010 to 0.10 incl.
Lead From from 0.0002 to 0.020 incl.
Silver From 0.0005 to 0.020 incl.
Antimony From 0,0010 to 0.050 incl.
Tellurium From 0.0005 to 0,010 incl.
Titan From 0,00010 to 0,0050 incl.
Chrome From 0,0003 to 0,030 incl.
Zinc From 0,0003 to 0,030 incl.

6.1 Accuracy (trueness and precision)

6.1.1 indicators of the accuracy of the method

Indicators of accuracy of the method: limit of the interval in which with a probability ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраof 0.95, is the absolute error of the results of the analysis (assigned error) ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, standard deviation of repeatability ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, the values of the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand reproducibility limit ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраdepending on the mass fraction of the element-impurities listed in table 6.


Table 6 — indicators of the accuracy of the method (ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра0,95)

Percentage

             
The level of the mass fraction of detectable elements

The boundaries of the interval of absolute error ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The standard deviation of repeatability ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The critical range
ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The standard deviation of intermediate precision
ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

The limit of intermediate precision
ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

Limit
of rehabilitation
of conductivity ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,00010 0,00007
0,00002 0,00008 0,00003 0,00009 0,00011
0,0003 0,0001
0,00007 0,0002 Of 0.00006 0,0002 0,0002
0,0005 0,0002
0,00014 0,0005 0,00008 0,0002 0,0003
0,0008 0,0003
0,00020 0,0007 0,00011 0,0003 0,0004
0,0010 0,0005
0,0003 0,0010 0,00018 0,0005 About 0.0006
0,0020 0,0010
0,0005 0,0019 0,00050 0,0014 0,0017
0,0030 0,0015
0,0008 0,0028 0,0007 0,0019 0,0022
0,0050 0,0022
0,0011 0,0040 0,0008 0,0022 0,0027
0,010 0,004
0,0019 0,007 0,0017 0,005 0,006
0,020 0,007
0,0035 0,013 0,0030 0,008 0,010
0,030 0,010
0,0050 0,018 0,0048 0,013 0,016
0,050 0,015
0,0072 0,026 0,0067 0,019 0,022
0,080 0,022
0,010 0,036 0,0090 0,025 0,030
0,10 0,03
0,013 0,05 0,012 0,03 0,04

6.1.2 Correctness

To estimate the systematic error of this method for the determination of impurity elements in palladium are used as reference values certified values of mass fractions of elements in the state standard samples of composition of GSO 7615−99 of palladium (set PD-36), GEO 7331−96 (set PD-28) or other GEO, the next set of detectable elements and metrological characteristics.

Systematic error of the method at the significance level ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра5% nonsignificant according to GOST R ISO 5725−4 for all defined elements-admixtures in palladium at all levels detectable concentrations.

6.1.3 Precision

6.1.3.1 Range of results four definitions obtained for the same sample by one operator using the same equipment within the shortest possible time intervals that may exceed the specified in table 6 critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраfor ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра4 according to GOST R ISO 5725−6 on average not more than once in 20 cases with proper use of the method.

6.1.3.2 within a single laboratory result analysis of the same sample obtained by different operators using the same equipment on different days, can vary with the excess specified in table 6 the limit of intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраaccording to GOST R ISO 5725−3 on average not more than once in 20 cases with proper use of the method.

6.1.3.3 the results of the analysis of the same samples obtained by two laboratories may vary with the excess specified in table 6 reproducibility limit ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраaccording to GOST R ISO 5725−1 on average not more than once in 20 cases with proper use of the method.

For intermediate values of the mass fraction values in terms of accuracy find using linear interpolation according to the formula (1) specified in 5.1.1.

6.2 Requirements

6.2.1 General requirements and safety requirements

General requirements for method of analysis and requirements for security of works and environmental safety — according to GOST R 52599.

6.2.2 qualifications of performers

The analysis allowed persons over the age of 18, trained in the prescribed manner and allowed to work independently on used equipment.

6.3 measurement Means, auxiliary devices, materials and reagents


Spectrometer with a generator of the arc is constant (AC) current or complex atomic-emission spectral analysis with the analyzer emission spectra of Maes type, designed to produce spectra in the range 210−350 nm.

Laboratory scales according to GOST 24104 with a limit of permissible absolute error of less than ±0.001 g.

Electrodes of graphite [1] 6 mm in diameter with a crater depth of 1−3 mm and a diameter of 4 mm.

Electrodes of graphite [1] 6 mm in diameter, sharpened to a truncated cone or hemisphere.

Electric stove with a closed spiral.

Glasses chemical resistant.

Hydrochloric acid OS.h. according to GOST 14261.

Rectified ethyl alcohol according to GOST 18300.

Distilled water according to GOST 6709.

Samples for calibration (samples of palladium with the previously established values of the mass fraction of impurity elements).

Standard samples of composition of palladium with an error of certified values mass fraction of impurities not exceeding 1/3 of the value of the absolute error of the present method for each level in mass fraction.

Allowed the use of other measurement means, auxiliary devices, materials and reagents, subject to obtaining accuracy rate that rivals those in table 6.

6.4 Sampling and sample preparation

6.4.1 Selection of laboratory samples from ingots or powder of refined palladium is carried out according to GOST R 52244.

6.4.2 Sample of palladium can enter the analysis in the form of ribbon, wire, shavings, sponge, powder.

6.4.3 Samples received for analysis in the form of tape, wire or shavings to remove the surface dirt is boiled in hydrochloric acid, diluted 1:1, for 10−15 min. the resulting solution was decanted, the sample was washed with distilled water by decantation 4−5 times and dried in air. Samples of the powder and sponge not treated with acid.

6.4.4 the laboratory From samples of palladium have selected 4 sample, from the samples to the calibration or standard samples — 2 samples weighing between 100 to 150 mg each, depending on depth electrodes used, but are the same for one spectrogram. The sample in powder form are pressed into the crater of the graphite electrode.

6.5 Preparation of equipment for measurement

6.5.1 Equipment ready to work according to the instructions. Operating modes of the spectrometer are given in table 7. The use of other operating modes subject to receipt of the accuracy rate that rivals those in table 6.


Table 7 — Recommended operating modes of the spectrometer

   
The name of the parameter
The value of the parameter
DC-arc:
 
the strength of the current, And
10−12
Alternating current arc:
 
the strength of the current, And
8−10
Conditions of registration of spectra:
 
gap width, mm
0,015
the analytical gap, mm
2,5
time roasting
5
the exhibition, with
15

6.5.2 electrode holders clear alcohol from surface contamination.

6.5.3 Include water cooling of the electrode holders.

6.5.4 Prepared for analysis a sample of palladium was placed in a graphite crater electrode. Contradictator is a graphite rod sharpened to a hemisphere or truncated cone. When used as a driving source of a DC arc of the analyzed sample is the anode.

6.5.5 the Interelectrode gap set on the magnified image of the arc on the screen, the intermediate diaphragm of 5 mm and maintain strictly constant, adjusting it throughout the exposure.

6.6 measurements

6.6.1 To obtain a calibration based on measurements taken of the intensity of the analytical lines of the determined elements and lines of comparison for the standard samples (calibration samples). For each of the identified elements choose one of the recommended analytical lines. The wavelengths of the analytical lines given in table 8. The use of other analytical lines subject to receipt of the accuracy rate that rivals those in table 6.


Table 8 — the wavelengths of the analytical lines

In nanometers

     
The name of the element Wavelength of analytical line
Wavelength comparison lines (palladium)
Aluminium 309,27 237,21
    307,52
Bismuth 289,79 Background
  306,77 Background
Iron 296,69 Background
  302,06 302,17; 307,52
  302,11 Background
Gold 267,59 268,62, background
Iridium 266,47 271,57, background
  292,48 307,52
  322,08 307,52
Calcium 315,88 Background
Cobalt 241,16 Background
  345,35  
Silicon 251,92 Background
  242,85 307,52
Magnesium 280,27 Background
Manganese 260,56 Background
  279,48 307,52
  259,37 Background
Copper 324,73 321,89, background
  296,1 Background
Nickel 341,47 332,09
  301,20 Background
  310,5 Background
  305,08 307,52
  339,29 307,52
Tin 317,50 Background
Osmium 290,90 237,21
Platinum 270,58 268,62
  299,79 Background
  265,94 307,52
Rhenium 342,46 237,21
Rhodium 332,30 362,09
  343,48 307,52, background
Ruthenium 366,13 356,66
  298,89 Background
  343,67 307,52
Lead 283,30 237,21
Silver 309,91 237,21
  338,29 Background
Antimony 259,80 Background
Tellurium 238,57 Background
Titan 308,80 237,21
  318,08 237,21
Chrome 284,32 Background
Zinc 334,50 Background
  328,23 Background
  Of 213.85 Of 213.85



The measurements were carried out for two batches of standard samples of composition of palladium or samples for the calibration and get the average.

6.6.2 Get the calibration based on relative or absolute intensity of the analytical lines of the determined elements from the mass fraction of these elements in standard sample (sample for graduation) to coordinate the program of the spectrometer.

Next, carry out the construction of the calibration characteristics for each element in the algorithm incorporated in the software of the spectrometer. Calibration characteristics stored in a data file of the calibration and then used for subsequent analyses.

6.6.3 Measured intensity of analytical lines of the determined elements for each of the four batches of samples.

6.6.4 using the calibration dependencies get four results of parallel measurements of the mass fraction of each analyzed element in the sample.

6.7 evaluation of the acceptability of the results of parallel measurements and obtaining the final result of the analysis

6.7.1 Acceptability of the results of parallel measurements was evaluated in accordance with GOST R ISO 5725−6 by mapping the range of these results ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраwith the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраshown in table 6.

6.7.2 If the range of the results of four parallel measurements ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраdoes not exceed the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, all the results and recognise acceptable for the final result of the analysis take the average of results of four parallel measurements.

6.7.3 If the range of the results of four parallel measurements exceeds ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, spend four parallel determinations.

Expect the critical range for up to eight parallel measurements ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраaccording to the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, (4)


where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis the standard deviation of repeatability.

If the received eight parallel measurements the value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраdoes not exceed the critical range ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, as a final result of the analysis taking the arithmetic mean of results of eight parallel measurements. Otherwise, as a final result of the analysis taking the median of the results of eight parallel measurements, if the regulations of the enterprise stipulate otherwise.

6.8 Control of accuracy of analysis results

6.8.1 Control of intermediate precision and reproducibility

When monitoring intermediate precision (variable factors of the operator and time) is the absolute difference of the two results of the analysis of the same sample obtained by different operators using the same equipment on different days, should not exceed the limit of intermediate precision ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраspecified in table 6.

Under the control of the reproducibility of the absolute disparity of the two results of the analysis of the same samples, obtained in two laboratories, in accordance with the requirements of this standard must not exceed the reproducibility limit R given in table 6.

6.8.2 verification of correct

The checking is carried out by analyzing standard samples of composition of palladium. The samples used for control of correctness cannot be used to obtain the calibration dependencies.

When checking the difference between the analysis result and the accepted reference (certified) value of the mass fraction of element-impurities in the standard sample should not exceed a critical value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

The critical value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis calculated by the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, (5)


where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — the error of setting the reference (certified) value of mass fraction of element-impurities in the standard sample;

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — the boundaries of the interval of the absolute error of the result of the analysis (values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраshown in table 6).

Annex a (informative). Table of values of lg (I (l)/I (f)), corresponding to the measured values of Delta S/"gamma»

Appendix A
(reference)


A table of values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраcorresponding to measured values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра


The following table A. 1 serves for transmission of measured values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

The table contains the results of the calculation in the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, (A. 1)


where ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis the difference in the density of pochernenija on the plate;

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра — contrast ratio.

We denote the total intensity of the line together with the background ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, the intensity of the background under the high line in the absence of the line ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

Since ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, the intensity ratio of the line ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраto the background intensity is determined by the formula

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра. (A. 2)


If the conditions of photographing the spectrum chosen so that the blackening of a line with a background ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand background in the absence of the line ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраare in the normal range,

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраwhere ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра. (A. 3)


Hence, using the expression ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра,

get ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

Table A. 1 covers the most important practices for analytical work values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраfrom 0.05 to 1.9.

The table consists of two parts: the part covering values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраfrom 0.05 to 0.99, and part of the covering values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраfrom 1.00 to 1.9.

In the first part of the table in the first column are the values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраwith two decimal places, the figures in the heads of others count from 0 to 9 indicate the third decimal place values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

For example, ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра0,537: find the first column value of 0.53 and in the graph with figure 7 to determine the corresponding value of the logarithm ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра0,388.

The second part of the table built in a similar way with the difference that the first column lists the values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраwith one decimal place and the numbers in the heads of other boxes represent the second after the decimal sign values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра.

For example, ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраof 1.36: find the first column value of 1.3 and the graph from figure 6 find the value of logarithm of ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра1,341.

For values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраsmaller than had 0.301, the value is ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраnegative the minus sign over the characteristic (±1…).

Since ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, the table can be used to find the value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраcorresponding to the values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраin any method of measurement.

If the contrast ratio ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраis not measured, instead of the values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраin the table apply values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра, use this table in a similar way. If the measured value ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand 0.674 in the first column find the value of 0.67 in column 4 determine the meaning of the logarithm of 0,571.

It should be noted that the found thus the value of 0,571 represents not ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраand ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра. The precision of the analysis by the method of «three standards» the fact is almost no effect.


Table A. 1 — Values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектраcorresponding to measured values ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

                     

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0 1 2 3 4 5 6 7 8 9
0,05

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,06

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,07

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,08

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,09

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,10

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,11

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,12

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,13

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,14

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,15

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,16

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,17

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,18

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,19

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,20

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,21

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,22

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,23

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,24

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,25

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,26

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,27

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,28

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

0,29

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

ГОСТ Р 52951-2008 Палладий. Методы атомно-эмиссионного анализа с дуговым возбуждением спектра

                     
0,30 1,998 0,000 0,002 0,004 0,006 0,008 0,010 0,012 0,014 0,016
0,31 0,018 0,020 0,022 0,024 0,026 0,028 0,029 0,031 0,033 0,035
0,32 0,037 0,039 0,041 0,043 0,045 0,047 0,049 0,050 0,052 0,054
0,33 0,056 0,058 0,060 0,062 0,064 0,065 0,067 To 0.069 0,071 0,073
0,34 0,075 0,077 0,078 0,080 0,082 0,084 0,086 0,088 0,089 0,091
0,35 0,093 0,095 0,097 0,098 0,100 Is 0.102 0,104 0,106 0,107 0,109
0,36 0,111 0,113 0,114 0,116 0,118 0,120 0,121 0,123 0,125 To 0.127
0,37 0,128 0,130 0,132 0,134 0,135 0,137 0,139 0,141 0,142 0,144
0,38 Of 0.146 0,147 0,149 0,151 0,153 0,154 0,156 0,158 0,159 0,161
0,39 0,163 0,164 0,166 0,168 0,170 0,171 0,173 0,175 0,176 0,178
0,40 0,180 Of 0, 181 0,183 0,184 0,186 0,188 0,189 0,191 0,193 0,194
0,41 0,196 0,198 0,199 0,201 0,203 0,204 0,206 0,207 0,209 0,211
0,42 Of 0.212 0,214 0,215 0,217 Of 0.219 0,220 0,222 0,224 0,225 0,227
0,43 0,228 0,230 0,231 0,233 0,235 Of 0.236 0,238 0,239 0,241 0,243
0,44 0,244 0,246 0,247 0.249 V. 0,250 0,252 0,253 0,255 Of 0.257 0,258
0,45 Is 0.260 0,261 Of 0.263 0,264 0,266 0,267 0,269 0,270 0,272 0,274
0,46 0,275 0,277 0,278 0,280 0,281 0,283 0,284 0,286 0,287 0,289
0,47 0,290 0,292 0,293 0,295 0,296 0,298 0,299 Had 0.301 0,302 0,304
0,48 0.305 per 0,307 0,308 0,310 0,311 0,313 0,314 0,316 0,317 0,319
0,49 0,320 0,322 0,323 0,325 0,326 0,328 0,329 0,331 0,332 0,333
0,50 0,335 0,336 0,338 0,339 0,341 0,342 0,344 0,345 0,347 0,348
0,51 0,349 0,351 0,352 0,354 0,355 0,357 0,358 0,360 0,361 0,362
0,52 0,364 0,365 0,367 0,368 0,370 0,371 0,372 0,374 0,375 0,377
0,53 0,378 0,380 0,381 0,382 0,384 0,385 0,387 0,388 0,389 0,391
0,54 0,392 0,394 0,395 0,396 0,398 0,399 0,401 0,402 0,403 0,405
0,55 0,406 0,408 0,409 0,410 0,412 0,413 0,415 0,416 0,417 0,419
0,56 0,420 0,421 0,423 0,424 0,426 0,427 0,428 0,430 0,431 0,432
0,57 0,434 0,435 0,437 0,438 0,439 0,441 0,442 0,443 0,445 0,446
0,58 0,447 0,449 0,450 0,452 0,453 0,454 0,456 0,457 0,458 0,460
0,59 0,461 0,462 0,464 0,465 0,466 0,468 0,469 0,470 0,472 0,473
0,60 0,474 0,476 0,477 0,478 0,480 0,481 0,482 0,484 Appears to be 0.485 0,486
0,61 To 0.488 0,489 0,490 0,492 0,493 0,494 0,496 0,497 0,498 0,500
0,62 0,501 0,502 0,504 0,505 To 0.506 0,507 0,509 0,510 0,511 0,513
0,63 To 0.514 0,515 0,517 0,518 0,519 0,521 0,522 0.523 accordingly 0,524 0,526
0,64 0,527 0,528 which 0,530 0,531 0,532 0,533 0,535 0,536 0,537 0,539
0,65 0,540 0,541 0,543 0,544 0,545 0,546 0,548 0,549 0,550 0,551
0,66 0,553 0,554 0,555 0,557 0,558 0,559 0,560 0,562 0,563 0,564
0,67 0,566 0,567 0,568 0,569 0,571 0,572 0,573 0,574 0,576 0,577
0,68 0,578 0,579 0,581 0,582 0,583 0,585 0,586 0,587 0,588 0,590
0,69 0,591 0,592 0,593 0,595 0,596 0,597 0,598 0,600 0,601 0,602
0,70 0,603 0,605 0,606 0,607 It was 0.608 0,610 0,611 0,612 0,613 0,615
0,71 0,616 0,617 0,618 0,620 0,621 0,622 0,623 0,624 0,626 0,627
0,72 0,628 0,629 0,631 0,632 0,633 0,634 0,636 0,637 0,638 0,639
0,73 0,641 0,642 0,643 0,644 0,645 0,647 0,648 0,649 0,650 0,652
0,74 0,653 0,654 0,655 0,656 0,658 0,659 0,660 0,661 0,663 0,664
0,75 0,665 0,666 0,667 0,669 0,670 0,671 0,672 0,673 0,675 0,676
0,76 0,677 Of 0.678 0,680 0,681 0,682 0,683 0,684 0,686 0,687 Of 0.688
0,77 0,689 0,690 0,692 0,693 0,694 0,695 0,696 0,698 0,699 0,700
0,78 0,701 0,702 0,704 0,705 0,706 0,707 0,708 0,710 0,711 0,712
0,79 0,713 0,714 0,716 0,717 0,718 0,719 0,720 0,721 0,723 0,724
0,80 0,725 0,726 0,727 0,729 0,730 0,731 Of 0.732 0,733 0,735 0,736
0,81 0,737 0,738 0,739 0,740 0,742 0,743 0,744 0,745 0,746 0,748
0,82 0,749 0.750 in 0,751 0,752 0,753 0,755 0,756 0,757 0,758 0,759
0,83 0,760 0,762 0,763 0,764 0,765 0,766 0,768 0,769 0,770 0,771
0,84 0,772 0,773 0,775 0,776 0,777 0,778 0,779 0,780 0,782 0,783
0,85 0,784 0,785 0,786 0,787 0,789 0,790 0,791 0,792 0,793 0,794
0,86 0,795 0,797 0,798 0,799 0,800 0,801 0,802 0,804 0,805 AZN 0.806
0,87 0,807 0,808 0,809 0,811 0,812 0,813 0,814 0,815 0,816 0,817
0,88 0,819 0,820 0,821 0,822 0,823 0,824 0,826 0,827 0,828 0,829
0,89 0,830 0,831 0,832 0,834 0,835 0,836 0,837 0,838 0,839 0,840
0,90 0,842 0,843 0,844 0,845 0,846 0,847 0,848 0,850 0,851 0,852
0,91 0,853 0,854 0,855 0,856 0,858 0,859 0,860 0,861 0,862 0,863
0,92 0,864 0,866 0,867 0,868 0,869 0,870 0,871 0,872 0,873 0,875
0,93 0,876 0,877 Of 0.878 0,879 0,880 0,881 0,883 0,884 0,885 0,886
0,94 0,887 0,888 0,889 0,890 0,892 0,893 0,894 0,895 0,896 0,897
0,95 Of 0, 898 0,899 0,901 Of 0.902 0,903 0,904 Of 0.905 0,906 0,907 0,908
0,96 0,910 0,911 0,912 0,913 0,914 0,915 0,916 0,917 0,919 0,920
0,97 0,921 Of 0.922 0,923 0,924 0,925 0,926 0,927 0,929 0,930 0,931
0,98 0,932 0,933 0,934 0,935 0,936 0,938 0,939 0,940 0,941 0,942
0,99 0,943 Gravity 0,944 0,945 0,946 0,948 0,949 0,950 0,951 0,952 0,953
1,0 0,954 Of 0.965 0,976 0,987 Is 0.998 1,009 1,020 1,031 1,042 1,053
1,1 1,064 1,075 1,086 1,097 1,107 1,118 1,129 1,140 1,150 1,161
1,2 1,172 1,182 1,193 1,204 1,214 1,225 1,235 1,246 1,257 1,267
1,3 1,278 1,288 1,299 1,309 1,320 1,330 1,341 1,351 1,362 1,372
1,4 1,382 1,393 1,403 1,414 1,424 1,434 1,445 1,455 1,465 1,476
1,5 1,486 1,496 1,507 1,517 1,527 1,538 1,548 1,558 1,568 1,579
1,6 1,589 1,599 1,609 1,620 1,630 1,640 1,650 1,661 1,671 1,681
1,7 1,691 1,701 1,712 1,722 1,732 1,742 1,752 1,763 1,773 1,783
1,8 1,793 1,803 1,813 1,824 1,834 1,844 1,854 1,864 1,874 1,884
1,9 1,894 1,905 1,915 1,925 1,935 1,945 1,955 1,965 1,975 1,986

Bibliography

   
[1] THE 3497−001−51046676−2001 Graphite electrodes for emission spectral analysis