By visiting this site, you accept the use of cookies. More about our cookie policy.

GOST 12226-80

GOST R ISO 15353-2014 GOST P 55080-2012 GOST R ISO 16962-2012 GOST R ISO 10153-2011 GOST R ISO 10280-2010 GOST R ISO 4940-2010 GOST R ISO 4943-2010 GOST R ISO 14284-2009 GOST R ISO 9686-2009 GOST R ISO 13899-2-2009 GOST 18895-97 GOST 12361-2002 GOST 12359-99 GOST 12358-2002 GOST 12351-2003 GOST 12345-2001 GOST 12344-88 GOST 12350-78 GOST 12354-81 GOST 12346-78 GOST 12353-78 GOST 12348-78 GOST 12363-79 GOST 12360-82 GOST 17051-82 GOST 12349-83 GOST 12357-84 GOST 12365-84 GOST 12364-84 STATE STANDARD P 51576-2000 GOST 29117-91 GOST 12347-77 GOST 12355-78 GOST 12362-79 GOST 12352-81 GOST P 50424-92 STATE STANDARD P 51056-97 GOST P 51927-2002 GOST P 51928-2002 GOST 12356-81 GOST R ISO 13898-1-2006 GOST R ISO 13898-3-2007 GOST R ISO 13898-4-2007 GOST R ISO 13898-2-2006 STATE STANDARD P 52521-2006 GOST P 52519-2006 GOST R 52520-2006 GOST P 52518-2006 GOST 1429.14-2004 GOST 24903-81 GOST 22662-77 GOST 6012-2011 GOST 25283-93 GOST 18318-94 GOST 29006-91 GOST 16412.4-91 GOST 16412.7-91 GOST 25280-90 GOST 2171-90 GOST 23401-90 GOST 30642-99 GOST 25698-98 GOST 30550-98 GOST 18898-89 GOST 26849-86 GOST 26876-86 GOST 26239.5-84 GOST 26239.7-84 GOST 26239.3-84 GOST 25599.4-83 GOST 12226-80 GOST 23402-78 GOST 1429.9-77 GOST 1429.3-77 GOST 1429.5-77 GOST 19014.3-73 GOST 19014.1-73 GOST 17235-71 GOST 16412.5-91 GOST 29012-91 GOST 26528-98 GOST 18897-98 GOST 26529-85 GOST 26614-85 GOST 26239.2-84 GOST 26239.0-84 GOST 26239.8-84 GOST 25947-83 GOST 25599.3-83 GOST 22864-83 GOST 25599.1-83 GOST 25849-83 GOST 25281-82 GOST 22397-77 GOST 1429.11-77 GOST 1429.1-77 GOST 1429.13-77 GOST 1429.7-77 GOST 1429.0-77 GOST 20018-74 GOST 18317-94 STATE STANDARD P 52950-2008 GOST P 52951-2008 GOST 32597-2013 GOST P 56307-2014 GOST 33731-2016 GOST 3845-2017 STATE STANDARD P ISO 17640-2016 GOST 33368-2015 GOST 10692-2015 GOST P 55934-2013 GOST P 55435-2013 STATE STANDARD P 54907-2012 GOST 3845-75 GOST 11706-78 GOST 12501-67 GOST 8695-75 GOST 17410-78 GOST 19040-81 GOST 27450-87 GOST 28800-90 GOST 3728-78 GOST 30432-96 GOST 8694-75 GOST R ISO 10543-99 GOST R ISO 10124-99 GOST R ISO 10332-99 GOST 10692-80 GOST R ISO 17637-2014 GOST P 56143-2014 GOST R ISO 16918-1-2013 GOST R ISO 14250-2013 GOST P 55724-2013 GOST R ISO 22826-2012 GOST P 55143-2012 GOST P 55142-2012 GOST R ISO 17642-2-2012 GOST R ISO 17641-2-2012 GOST P 54566-2011 GOST 26877-2008 GOST R ISO 17641-1-2011 GOST R ISO 9016-2011 GOST R ISO 17642-1-2011 STATE STANDARD P 54790-2011 GOST P 54569-2011 GOST P 54570-2011 STATE STANDARD P 54153-2010 GOST R ISO 5178-2010 GOST R ISO 15792-2-2010 GOST R ISO 15792-3-2010 GOST P 53845-2010 GOST R ISO 4967-2009 GOST 6032-89 GOST 6032-2003 GOST 7566-94 GOST 27809-95 GOST 22974.9-96 GOST 22974.8-96 GOST 22974.7-96 GOST 22974.6-96 GOST 22974.5-96 GOST 22974.4-96 GOST 22974.3-96 GOST 22974.2-96 GOST 22974.1-96 GOST 22974.13-96 GOST 22974.12-96 GOST 22974.11-96 GOST 22974.10-96 GOST 22974.0-96 GOST 21639.9-93 GOST 21639.8-93 GOST 21639.7-93 GOST 21639.6-93 GOST 21639.5-93 GOST 21639.4-93 GOST 21639.3-93 GOST 21639.2-93 GOST 21639.0-93 GOST 12502-67 GOST 11878-66 GOST 1763-68 GOST 13585-68 GOST 16971-71 GOST 21639.10-76 GOST 2604.1-77 GOST 11930.7-79 GOST 23870-79 GOST 11930.12-79 GOST 24167-80 GOST 25536-82 GOST 22536.2-87 GOST 22536.11-87 GOST 22536.6-88 GOST 22536.10-88 GOST 17745-90 GOST 26877-91 GOST 8233-56 GOST 1778-70 GOST 10243-75 GOST 20487-75 GOST 12503-75 GOST 21548-76 GOST 21639.11-76 GOST 2604.8-77 GOST 23055-78 GOST 23046-78 GOST 11930.11-79 GOST 11930.1-79 GOST 11930.10-79 GOST 24715-81 GOST 5639-82 GOST 25225-82 GOST 2604.11-85 GOST 2604.4-87 GOST 22536.5-87 GOST 22536.7-88 GOST 6130-71 GOST 23240-78 GOST 3242-79 GOST 11930.3-79 GOST 11930.5-79 GOST 11930.9-79 GOST 11930.2-79 GOST 11930.0-79 GOST 23904-79 GOST 11930.6-79 GOST 7565-81 GOST 7122-81 GOST 2604.3-83 GOST 2604.5-84 GOST 26389-84 GOST 2604.7-84 GOST 28830-90 GOST 21639.1-90 GOST 5640-68 GOST 5657-69 GOST 20485-75 GOST 21549-76 GOST 21547-76 GOST 2604.6-77 GOST 22838-77 GOST 2604.10-77 GOST 11930.4-79 GOST 11930.8-79 GOST 2604.9-83 GOST 26388-84 GOST 14782-86 GOST 2604.2-86 GOST 21639.12-87 GOST 22536.8-87 GOST 22536.0-87 GOST 22536.3-88 GOST 22536.12-88 GOST 22536.9-88 GOST 22536.14-88 GOST 22536.4-88 GOST 22974.14-90 GOST 23338-91 GOST 2604.13-82 GOST 2604.14-82 GOST 22536.1-88 GOST 28277-89 GOST 16773-2003 GOST 7512-82 GOST 6996-66 GOST 12635-67 GOST 12637-67 GOST 12636-67 GOST 24648-90

GOST 12226−80 Platinum. Methods of analysis (modified N 1, 2, 3, 4, 5)


GOST 12226−80

Group B59


INTERSTATE STANDARD


PLATINUM

Methods of analysis

Platinum. Methods of analysis

AXTU 1709

Date of introduction 1981−07−01


INFORMATION DATA

1. DEVELOPED by the Ministry of nonferrous metallurgy of the USSR

DEVELOPERS

V. A. Korneev, V. E. Avramov, M. A. Gavrilov, G. N. Verkhoturov, A. K., Dementieva, I., Karshakevich

2. APPROVED AND put INTO EFFECT by Decision of the USSR State Committee for standards from 18.03.80 N 1201

Change No. 5 adopted by the Interstate Council for standardization, Metrology and certification (Protocol No. 10 dated 04.10.96)

Was the Technical Secretariat of MGS N 2251

The adoption voted:

   
The name of the state
The name of the national authority for standardization
The Republic Of Azerbaijan
Azgosstandart
The Republic Of Belarus
Gosstandart Of Belarus
The Republic Of Georgia
Gosstandart
The Republic Of Kazakhstan
Gosstandart Of The Republic Of Kazakhstan
The Republic Of Moldova
Moldovastandart
Russian Federation
Gosstandart Of Russia
The Republic Of Tajikistan
Tajikistandart
Turkmenistan
The main state inspection of Turkmenistan
The Republic Of Uzbekistan
Standards

3. The frequency of inspection — 5 years

4. REPLACE GOST 12226−66

5. REFERENCE NORMATIVE AND TECHNICAL DOCUMENTS

   
The designation of the reference document referenced Item number
GOST 5962−67*
2.1, 3.1
GOST 6563−75
4.1
GOST 9147−80
4.1
GOST 10691.0−84
3.1
GOST 10691.1−84
3.1
GOST 14261−77 2.1, 3.1
GOST 22864−83
1.1
GOST 25336−82 4.1

________________
* GOST 5962−67 not valid on the territory of the Russian Federation from 01.07.2001. Valid GOST R 51652−2000. «Rectified ethyl alcohol from food raw material. Technical conditions». Note the «CODE"

6. The expiration time limit is removed by the Resolution of Gosstandart from 24.02.92 N 174

7. REVISED (December 1998), with Changes N 1, 2, 3, 4, 5, approved in July 1985, March 1989, December 1990, February 1992, February 1997 (IUS 10−85, 6−89, 4−91, 6−92, 5−97).


This standard establishes spectrometry and spectrographic methods for determining the content of palladium, iridium, rhodium, ruthenium, gold, silver, iron, Nickel, copper, lead, silicon, zinc, aluminum, tin, antimony, magnesium and gravimetric method for the determination of volatile impurities.

The spectrometric method is based on spark excitation radiation between the sample of platinum and contradictator or translation of a sample of platinum in the globule, the evaporation of the elements in the arc of DC with subsequent photoelectric measurement of intensity of analytical lines of impurities.

Determine impurities: palladium, rhodium, iridium, ruthenium, lead — from 0.002 to 0.1 percent; magnesium, silicon, copper, aluminum, from 0.001 to 0.02% each; gold, Nickel, iron, zinc, silver, from 0.001 to 0.1% each; antimony — from 0.0005% to 0.02%; and tin from 0.0005 to 0.1%.

Spectrographic method is based on the translation of the shavings or powder of platinum in the globule, the evaporation of the elements of the impurities in the arc of DC and photographic registration of the spectra.

Determine impurities: palladium from 0.002 to 0.1%; iridium, rhodium, ruthenium is from 0.005 to 0.1% each; copper, Nickel, gold, magnesium, silver, from 0.001 to 0.01% each; aluminum, silicon, lead, zinc, from 0.001 to 0.02% each; iron is from 0.002 to 0.05%; tin and antimony from 0.0005 to 0.02% of each.

The gravimetric method is based on determining the loss on ignition by the mass difference of sample before and after calcination.

(Changed edition, Rev. N 2, 4, 5).

1. GENERAL REQUIREMENTS

1.1. General requirements for methods of analysis GOST 22864.

(Changed edition, Rev. N 3).

1.2. (Deleted, Rev. N 3).

1.3. Mass fraction of impurities spectrographic and gravimetric methods determine at least four parallel batches; MS — at least four seats is one pill.

1.4,1.5. (Deleted, Rev. N 3).

2. SPECTROMETRIC METHOD

2.1. Apparatus, reagents and solutions

Quantometer emission ARL 31000 or other device of equal accuracy.

The generator of the unipolar low-voltage spark.

Press the str-60.

Steel mold with a matrix of an inner diameter of 40 mm.

A machine for sharpening metal bars.

Analytical scale.

Platinum rods with a diameter of 6 mm, length 150 mm (mass fraction of platinum 99,98%), sharpened on a cone at an angle of 90°.

The calibration samples.

Rectified ethyl alcohol according to GOST 5962.

Hydrochloric acid of high purity according to GOST 14261, diluted 1:1.

Distilled water.

Standard sample of platinum to verify the correctness of analysis results.

The spectral graphite electrodes net.

(Changed edition, Rev. N 3, 5).

2.2. Preparation for assay

For removal of surface contamination of platinum is boiled in hydrochloric acid for 2 min, washed with water and dried.

A sample of 30 g was placed in a mold and pressed into a tablet with a force of 540,000 N.

The tablet is placed on a stand water-cooled air tripod spectrometer. Contradictator is a platinum rod. Analytical gap 3±0.02 mm is installed by the template.

Tablet and a platinum rod electrodes are unipolar low-voltage spark.

If you come to the analysis of samples in the form of cast rods with a diameter of 6−8 mm, length 20−25 mm, the ends of the rods processed on a flat smooth surface, after which the platinum surface is cleaned as stated above.

The analysis in the arc of a DC sample of platinum weighing 150 mg were placed in the crater of the graphite electrode with a diameter of 6 mm (depth of crater, 1.5−2.0 mm, diameter 3.5−4.0 mm). Contradictator serve graphite rods with a length of 30−50 mm, sharpened to a truncated cone and a platform with a diameter of 2.5 mm.

2.3. Analysis

Sample of platinum serves as the anode (+) and contradicted — cathode (-).

Preparation of equipment for operation is carried out according to the instructions eksluatatsii devices.

Sample obygryvaet at least four times. After each sparking for a given analytical program automatically prints the results of the measurement for each item.

Contradicted replaced with a new one before burning a new sample.

2.2, 2.3 (Revised edition, Edit. N 5).

2.4. Processing of the results

The results of measurements using a DC line built for the calibration specimens, determine the mass percent of impurities.

When equipped with quantometer computer (PC) according to the given analytical program to automatically conduct the calculation the mass fraction and printing.

The final result of the analysis be the arithmetic mean of four measurements (obygryvanii), the maximum difference between them does not exceed the allowable discrepancy at a confidence probability P = 0,95.

If carried out calibration of measuring channels of the spectrometer for the calibration samples, the mass fraction of elements, impurities obtained by multiplying the registered value of the readings of the digital voltmeter to a scale which for each element is determined during calibration.

Analytical lines for analysis are presented in table.1.

Table 1

   
The designated element
Wavelength, km
Palladium
355,30; 340,45
Rhodium
365,79; 343,48
Iridium
269,42; 224,26
Ruthenium
349,89; 240,27
Gold
267,59
Lead
405,78
Iron
296,68; 259,94
Silicon
251,61; 288,15
Tin
286,33; 189,9
Aluminium
396,15
Antimony
206,83; 231,14
Silver
328,06; 338,28
Copper
324,75; 327,39
Nickel
227,02; 361,93; 221,61
Magnesium
279,55; 285,21
Zinc
334,50; of 213.85
Platinum
271,31; 280,85 Internal standard



(Changed edition, Rev. N 3, 4).

2.4.1. The preparation of calibration samples (see section 3.4.3).

2.4.2. Allowable absolute differences of the results of the parallel definitions should not exceed the values given in table.3.

3. SPECTROGRAPHIC METHOD

3.1. Apparatus, reagents and solutions

A diffraction spectrograph with a grating of 600 lines/mm.

Three-step attenuator with transmittance levels of 100, 40 and 10%.

Generator DC arc power up to 15 A.

Geregistreerde microphotometer.

Steel mold with a matrix of an inner diameter of 4 mm.

Grinding machine graphite electrodes.

The spectral graphite electrodes-pure grades of high purity 7−3, high purity 7−4, b-3, with a diameter of 6 mm.

The calibration samples.

Photographic plates of the spectral type of ES sensitivity of 5−10 units GOST 10691.1 or type 2 sensitivity 16 units

Rectified ethyl alcohol according to GOST 5962.

Hydrochloric acid of high purity according to GOST 14261, diluted 1:1.

Developer and fixer according to GOST 10691.0, GOST 10691.1. Allowed to use other contrast working developer.

Standard sample of platinum to verify the correctness of analysis results.

(Changed edition, Rev. N 3).

3.2. Preparation for assay

For removal of surface contamination of platinum is boiled in hydrochloric acid for 2 min, washed with water and dried.

Sample weighing 150 mg was placed in a cratered graphite electrode (crater depth of 1.5−2.0 mm, diameter 4 mm). In the case of the analysis of the spongy platinum sample metal pressed with a force of 500−1000 N.

Contradictator serve graphite rods with a length of 30−50 mm, sharpened on pecheny cone (with a platform 2.5 mm diameter).

3.3. Analysis

The spectra of the calibration samples are photographed and analyzed on the spectrograph: the slit width of 0.015 mm, the light slit being a capacitor, the current of 10−12 A, exposition 60 s. the test sample is the anode. Analytical gap 2.5 mm adjust in the process of exhibiting spectra in the image on the intermediate diaphragm. From each sample of the calibration samples and specimens receive one spectrum.

Photographic plates showing for 5 min at developer temperature of 20 °C. Shown photographic plates rinsed in water, fixed, washed in running water, dried and photometric.

3.4. Processing of the results

3.4.1. For all elements of the internal standard serve line platinum. Determination of mass fraction of impurities is carried out by the method of «three standards».

Analytical lines for analysis are presented in table.2.

Table 2

     

The designated element
Wavelength, nm
  analytical lines
internal standard
Palladium
325,87
326,84
  355,30
366,30
Rhodium
339,68
326,84
  365,79
366,30
Iridium
269,42
271,57
  284,97
271,57
Ruthenium
287,49
280,85
  343,67
326,84
Gold
267,59
267,45
  312,27
311,98
Silver
328,06
326,84
Nickel
303,79
307,95
  341,47
348,34
Copper
327,39
326,84
Iron
302,06
302,43
  259,93
259,60
Aluminium
309,27
307,95
Silicon
288,15
285,31
Tin
283,99
285,31
Zinc
334,50
326,84
Lead
280,20
285,31
Antimony
259,80
259,60
Magnesium
280,26
285,31
  285,21
285,31

3.4.2. Using microphotometry measure the amount of blackening of analytical lines and the internal standard.

Using the characteristic curve, which is constructed for each spectrogram, determine the logarithm of the ratio of the intensity of analytical lines of impurities and the internal standard.
The calibration graphs are built in coordinates ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5), x — axis is the ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5)(logarithm of the mass fraction of the calibration samples); on the y-axisГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5)(the logarithm of the ratio of the intensity of the impurity line to the line of the internal standard).

In the region of high mass fraction allowed the construction of calibration graphs in the coordinate ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5), where ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5)is the difference of pochernenija analytical lines and the internal standard.

For the calibration chart to find the mass fraction of impurities in the analyzed samples.

The final result of the analysis be the arithmetic mean of four parallel measurements, the maximum difference between them does not exceed the permissible discrepancy at a confidence probability P = 0,95.

3.4.3. The preparation of calibration samples

Calibration samples for spectral analysis were prepared by direct introduction of the calculated quantity of powders of impurities (grade hç) in the spectral pure platinum (mass fraction of platinum is not less than 99,98%).

Sample impurities with a small amount of platinum pressed into a tablet (to avoid loss of impurities), then the pill and the rest of platinum is placed in a graphite crucible and melted in a heating furnace of the type IST-016 at 1950 °C for 20 min.

After melting the ingots are purified from mechanical inclusions, washed in hydrochloric acid and rasstraivaet chips of size 1 mm. Chips boil in hydrochloric acid, washed with distilled water and dried.

The purity of platinum up to determine the melting spectrographic method under the same conditions under which to conduct analysis. Impurities that are found in platinum, is determined by a method of additives, and found the mass fraction take into account when producing the calibration samples.

Prepare a series of six calibration samples in the range of mass fraction of from 0.0005 to 0.1%. Allowed the preparation of calibration samples by another method ensuring the specified accuracy of the analysis.

3.4.4. The difference between the results of parallel measurements and the results of the analysis shall not exceed the values of permissible differences given in table.3.

Table 3

     

Mass fraction of impurities, %
Allowable difference, %
  parallel definitions
the results of the analysis
To from 0,0005 0,0010 incl.
0,0010
0,0010
SV. 0,001 «0,003 « 0,004 0,005
» 0,003 «0,010 « 0,006 0,008
» Is 0.01 «to 0.03 «
0,01
0,02
» To 0.03 «to 0.10 «
0,02 0,03

3.4.1−3.4.4. (Changed edition, Rev. N 3).

4. GRAVIMETRIC METHOD FOR THE DETERMINATION OF LOSS ON IGNITION

4.1. Equipment

Furnace for calcination at 800−900 °C.

Platinum crucibles N 100−4 according to GOST 6563.

Crucibles porcelain low N 3 according to GOST 9147.

Analytical scale.

Desiccator, according to GOST 25336.

(Changed edition, Rev. N 2).

4.2. Analysis

Clean platinum crucible with a constant specific mass is placed in a porcelain crucible and calcined in a furnace for 10−15 minutes at a temperature of 650−800 °C. after calcination porcelain crucible with platinum are removed from the oven, cooled in a desiccator to a temperature of (20±5) °C and determine the mass of the platinum crucible.

After that it is weighed 5.0 g of the analyzed metal. The crucible with metal and placed in a porcelain crucible and calcined in a furnace under the same conditions for 20−25 min, remove the porcelain crucible with the platinum crucible from the furnace and cooled in a desiccator to a temperature of (20±5) °C.

A platinum crucible is weighed, placed in a porcelain crucible and calcined under the same conditions for 10−15 min, then cooled in a desiccator and re-weighed. The operation of calcination and cooling, the crucible is repeated until obtaining constant weight of the platinum crucible with metal.

The difference between the mass of the crucible with the metal before and after incineration gives the mass loss during annealing in the sample.

(Changed edition, Rev. N 1, 2, 4).

4.3. Processing of the results

Mass fraction loss during annealing (aГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5)) percentage calculated by the formula

ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5)



where ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5)the difference in mass of the crucible with the metal before calcination and after calcination, g;

ГОСТ 12226-80 Платина. Методы анализа (с Изменениями N 1, 2, 3, 4, 5) — weight of metal,

For the results analysis be the arithmetic mean of four parallel measurements, the maximum difference between them does not exceed 0,005% at a mass fraction loss during annealing up to 0.02% and 0.01% when the mass fraction of loss on ignition of from 0.02 to 0.05% at confidence probability P = 0,95.

(Changed edition, Rev. N 4).