By visiting this site, you accept the use of cookies. More about our cookie policy.

GOST R 51056-97

GOST R ISO 15353-2014 GOST P 55080-2012 GOST R ISO 16962-2012 GOST R ISO 10153-2011 GOST R ISO 10280-2010 GOST R ISO 4940-2010 GOST R ISO 4943-2010 GOST R ISO 14284-2009 GOST R ISO 9686-2009 GOST R ISO 13899-2-2009 GOST 18895-97 GOST 12361-2002 GOST 12359-99 GOST 12358-2002 GOST 12351-2003 GOST 12345-2001 GOST 12344-88 GOST 12350-78 GOST 12354-81 GOST 12346-78 GOST 12353-78 GOST 12348-78 GOST 12363-79 GOST 12360-82 GOST 17051-82 GOST 12349-83 GOST 12357-84 GOST 12365-84 GOST 12364-84 STATE STANDARD P 51576-2000 GOST 29117-91 GOST 12347-77 GOST 12355-78 GOST 12362-79 GOST 12352-81 GOST P 50424-92 STATE STANDARD P 51056-97 GOST P 51927-2002 GOST P 51928-2002 GOST 12356-81 GOST R ISO 13898-1-2006 GOST R ISO 13898-3-2007 GOST R ISO 13898-4-2007 GOST R ISO 13898-2-2006 STATE STANDARD P 52521-2006 GOST P 52519-2006 GOST R 52520-2006 GOST P 52518-2006 GOST 1429.14-2004 GOST 24903-81 GOST 22662-77 GOST 6012-2011 GOST 25283-93 GOST 18318-94 GOST 29006-91 GOST 16412.4-91 GOST 16412.7-91 GOST 25280-90 GOST 2171-90 GOST 23401-90 GOST 30642-99 GOST 25698-98 GOST 30550-98 GOST 18898-89 GOST 26849-86 GOST 26876-86 GOST 26239.5-84 GOST 26239.7-84 GOST 26239.3-84 GOST 25599.4-83 GOST 12226-80 GOST 23402-78 GOST 1429.9-77 GOST 1429.3-77 GOST 1429.5-77 GOST 19014.3-73 GOST 19014.1-73 GOST 17235-71 GOST 16412.5-91 GOST 29012-91 GOST 26528-98 GOST 18897-98 GOST 26529-85 GOST 26614-85 GOST 26239.2-84 GOST 26239.0-84 GOST 26239.8-84 GOST 25947-83 GOST 25599.3-83 GOST 22864-83 GOST 25599.1-83 GOST 25849-83 GOST 25281-82 GOST 22397-77 GOST 1429.11-77 GOST 1429.1-77 GOST 1429.13-77 GOST 1429.7-77 GOST 1429.0-77 GOST 20018-74 GOST 18317-94 STATE STANDARD P 52950-2008 GOST P 52951-2008 GOST 32597-2013 GOST P 56307-2014 GOST 33731-2016 GOST 3845-2017 STATE STANDARD P ISO 17640-2016 GOST 33368-2015 GOST 10692-2015 GOST P 55934-2013 GOST P 55435-2013 STATE STANDARD P 54907-2012 GOST 3845-75 GOST 11706-78 GOST 12501-67 GOST 8695-75 GOST 17410-78 GOST 19040-81 GOST 27450-87 GOST 28800-90 GOST 3728-78 GOST 30432-96 GOST 8694-75 GOST R ISO 10543-99 GOST R ISO 10124-99 GOST R ISO 10332-99 GOST 10692-80 GOST R ISO 17637-2014 GOST P 56143-2014 GOST R ISO 16918-1-2013 GOST R ISO 14250-2013 GOST P 55724-2013 GOST R ISO 22826-2012 GOST P 55143-2012 GOST P 55142-2012 GOST R ISO 17642-2-2012 GOST R ISO 17641-2-2012 GOST P 54566-2011 GOST 26877-2008 GOST R ISO 17641-1-2011 GOST R ISO 9016-2011 GOST R ISO 17642-1-2011 STATE STANDARD P 54790-2011 GOST P 54569-2011 GOST P 54570-2011 STATE STANDARD P 54153-2010 GOST R ISO 5178-2010 GOST R ISO 15792-2-2010 GOST R ISO 15792-3-2010 GOST P 53845-2010 GOST R ISO 4967-2009 GOST 6032-89 GOST 6032-2003 GOST 7566-94 GOST 27809-95 GOST 22974.9-96 GOST 22974.8-96 GOST 22974.7-96 GOST 22974.6-96 GOST 22974.5-96 GOST 22974.4-96 GOST 22974.3-96 GOST 22974.2-96 GOST 22974.1-96 GOST 22974.13-96 GOST 22974.12-96 GOST 22974.11-96 GOST 22974.10-96 GOST 22974.0-96 GOST 21639.9-93 GOST 21639.8-93 GOST 21639.7-93 GOST 21639.6-93 GOST 21639.5-93 GOST 21639.4-93 GOST 21639.3-93 GOST 21639.2-93 GOST 21639.0-93 GOST 12502-67 GOST 11878-66 GOST 1763-68 GOST 13585-68 GOST 16971-71 GOST 21639.10-76 GOST 2604.1-77 GOST 11930.7-79 GOST 23870-79 GOST 11930.12-79 GOST 24167-80 GOST 25536-82 GOST 22536.2-87 GOST 22536.11-87 GOST 22536.6-88 GOST 22536.10-88 GOST 17745-90 GOST 26877-91 GOST 8233-56 GOST 1778-70 GOST 10243-75 GOST 20487-75 GOST 12503-75 GOST 21548-76 GOST 21639.11-76 GOST 2604.8-77 GOST 23055-78 GOST 23046-78 GOST 11930.11-79 GOST 11930.1-79 GOST 11930.10-79 GOST 24715-81 GOST 5639-82 GOST 25225-82 GOST 2604.11-85 GOST 2604.4-87 GOST 22536.5-87 GOST 22536.7-88 GOST 6130-71 GOST 23240-78 GOST 3242-79 GOST 11930.3-79 GOST 11930.5-79 GOST 11930.9-79 GOST 11930.2-79 GOST 11930.0-79 GOST 23904-79 GOST 11930.6-79 GOST 7565-81 GOST 7122-81 GOST 2604.3-83 GOST 2604.5-84 GOST 26389-84 GOST 2604.7-84 GOST 28830-90 GOST 21639.1-90 GOST 5640-68 GOST 5657-69 GOST 20485-75 GOST 21549-76 GOST 21547-76 GOST 2604.6-77 GOST 22838-77 GOST 2604.10-77 GOST 11930.4-79 GOST 11930.8-79 GOST 2604.9-83 GOST 26388-84 GOST 14782-86 GOST 2604.2-86 GOST 21639.12-87 GOST 22536.8-87 GOST 22536.0-87 GOST 22536.3-88 GOST 22536.12-88 GOST 22536.9-88 GOST 22536.14-88 GOST 22536.4-88 GOST 22974.14-90 GOST 23338-91 GOST 2604.13-82 GOST 2604.14-82 GOST 22536.1-88 GOST 28277-89 GOST 16773-2003 GOST 7512-82 GOST 6996-66 GOST 12635-67 GOST 12637-67 GOST 12636-67 GOST 24648-90

GOST R 51056−97 Steel alloyed and high alloy. Atomic emission spectral method for the determination of tungsten and molybdenum


GOST R 51056−97

Group B39


STATE STANDARD OF THE RUSSIAN FEDERATION

STEEL ALLOYED AND HIGH-ALLOYED

Atomic emission spectral method for the determination of tungsten
and molybdenum

Alloyed and high-alloyed steels.
Atomic emission spectrometry method for determination of tungsten
and molybdenum


OKS 77.040*
AXTU 0709
____________________
* In the index «National standards», 2008
OKS 77.080.20. — Note the manufacturer’s database.

Date of introduction 1998−01−01


Preface

1 DEVELOPED AND SUBMITTED by the Technical Committee TC 145 «monitoring Methods of steel products"

2 ADOPTED AND put INTO EFFECT by the Resolution of Gosstandart of Russia of June 2, 1997 No. 203

3 INTRODUCED FOR THE FIRST TIME

1 SCOPE


This standard specifies atomic emission (with excitation in inductively coupled plasma) method for determination of molybdenum and tungsten in alloyed and high-alloyed steels in the range of 0.01−5.0%.

The method is based on measuring the intensity of emission of atoms of tungsten and molybdenum with the introduction of the sample solution into an excitation source.

2 NORMATIVE REFERENCES


The present standard features references to the following standards:

GOST 3118−77 hydrochloric Acid. Specifications

GOST 4204−77 sulfuric Acid. Specifications

GOST 4461−77 nitric Acid. Specifications

GOST 6552−80 orthophosphoric Acid. Specifications

GOST 6709−72 distilled Water. Specifications

GOST 10157−79 Argon gaseous and liquid. Specifications

GOST 12349−83 Steel alloyed and high alloy. Methods for determination of tungsten

GOST 13610−79 carbonyl Iron radio. Specifications

GOST 18289−78 Sodium volframovich 2 water. Specifications

GOST 28473−90 Iron, steel, ferroalloys, chromium and manganese metal. General requirements for methods of analysis

3 GENERAL REQUIREMENTS


General requirements for method of analysis according to GOST 28473.

4 EQUIPMENT, REAGENTS, SOLUTIONS


Standard spectrometric system consisting of a spectrometer (multi-channel or sequential scanning), tripod excitation, high-frequency generator, the measuring electronic system and computer.

Argon according to GOST 10157.

Hydrochloric acid according to GOST 3118 and a solution of 1:9.

Nitric acid according to GOST 4461 and a solution of 1:1.

Orthophosphoric acid according to GOST 6552.

Sulfuric acid according to GOST 4204, and a solution of 1:1.

Distilled water according to GOST 6709.

A mixture of acids: 300 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof distilled water, 150 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof phosphoric acid, 150 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof nitric acid and 300 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof hydrochloric acid stir gently.

Carbonyl iron according to GOST 13610.

Molybdenum metal brand MCIT.

Standard solutions of molybdenum:

Solution N 1. 1 g of metallic molybdenum is dissolved in 20 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof nitric acid (1:1) and 20 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof sulphuric acid (1:1). The solution was evaporated to fumes of sulfuric acid, add 100 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof water and heated to dissolve the salts. After cooling, the solution is transferred to a volumetric flask with a capacity of 1 DMГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена, adjusted to the mark with water and mix. Standard solution 1 contains N 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена1 mg of molybdenum.

A Solution Of N 2. 10 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаstandard solution N 1 is transferred to a volumetric flask with a capacity of 100 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена, adjusted to the mark with water and mix. Standard solution No. 2 contains a 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена100 mcg of molybdenum.

Sodium volframovich 2-water according to GOST 18289.

Standard solutions of tungsten:

Solution N 3. 1,7942 volframovich g of sodium is dissolved in a volumetric flask with a capacity of 1 DMГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаin a small amount of water, bring to the mark with water and mix. Standard solution No. 3 contains a 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена1 mg of tungsten. The titer of the obtained solution set by the gravimetric method according to GOST 12349.

A Solution Of N 4. 10 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof the solution No. 3 is transferred to a volumetric flask with a capacity of 100 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена, adjusted to the mark with water and mix. Standard solution No. 4 contains 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена100 micrograms of tungsten.

A Solution Of N 5. In a volumetric flask with a capacity of 100 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаmaking shown in table 1 are the volumes of standard solutions N 1 and 3, hydrochloric acid solution (1:9) bring to mark and mix. Standard solution No. 5 contains of 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаto 10 micrograms of tungsten and molybdenum.


Table 1 — Composition of solutions for monitoring the stability of calibration parameters

           
Regulirovanie solutions
The ranges of determination of tungsten and molybdenum, %
Standard solutions

Mass fraction of tungsten
and molybdenum in recalibrating solutions
    N

Volume, cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена

µg/cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена

%
A and standard solution N 5 From 0.01 to 1.0 incl.
1
1,0
10,0
Of 1.00
    3
1,0
10,0 Of 1.00
B
From 0.01 to 1.0 incl.
2
0,5
0,5
0,050
    4 0,5 0,5 0,050
In and standard solution N 5 From 1.0 to 5.0 incl.
1
1,0
10,0
10,0
    3
1,0 10,0 10,0
G
From 1.0 to 5.0 incl.
2
0,5
0,5
0,50
    4
0,5 0,5 0,50

5 PREPARING FOR ANALYSIS

5.1 preparing the instrument for measurements carried out in accordance with the instructions for its use and maintenance. Instrumental parameters of the device and flows of argon is set in the range that the maximum sensitivity of the determination of tungsten and molybdenum.

5.2 establishing the calibration parameters is performed according to the solution and 5 N hydrochloric acid solution (1:9). For each solution perform at least five measurements of the intensities of tungsten (at the analytical line wavelength 207,91 or 218,94 nm) and molybdenum (analytical line wavelength 202,03 or 203,84 nm). According to the average values of the intensities of the calculated parameters calibration parameters. When determining the calibration characteristics of the mass fraction of tungsten and molybdenum is expressed as a percentage in accordance with the values given in table 1. Parameters calibration parameters recorded in the computer memory.

The use of other analytical lines, if they provide the definition of a mass fraction of tungsten and molybdenum in the required range with the accuracy stipulated in this standard.

The use of other methods of determining the parameters of the calibration parameters, if available mathematical software spectrometer setup.

5.3 the analysis of the steels with a mass fraction of tungsten and molybdenum, from 0.01 to 1.0% suspension of the sample mass 0,1000 g placed in a beaker with a capacity of 150 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена, 30 cm addГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаa freshly prepared mixture of acids, cover with a watch glass and heated until complete dissolution of the sample. The solution was cooled, washed the walls of the beaker and watch glass with distilled water, quantitatively transferred to a volumetric flask with a capacity of 100 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена, adjusted with distilled water to the mark and mix. The resulting solution contains 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена1 mg sample of the sample steel.

In the analysis of the steels with a mass fraction of tungsten and molybdenum in the range from 1.0 to 5.0% in a volumetric flask with a capacity of 50 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаadd 5 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof the resulting solution, adjusted with hydrochloric acid (1:9) to the mark and mix. The thus obtained solution contains 100 µg of sample sample 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена.

Allow the use of other methods of dissolution of batches of steels, providing a complete decomposition of the sample and does not require changes in further stages of analysis

.

5.4 dissolution of the sample state standard samples (GSO), similar in chemical composition to the analysed samples, carried out in accordance with 5.3.

5.5 in Parallel with the dissolution of the sample, performing all operations of the methods and using the same quantities of reagents, carry out a test to based on the content of tungsten and molybdenum in the reagents.

5.6 Preparation of solutions used to control the stability of the calibration parameters

5.6.1 Two hinges of iron, weighing 0.09 g each, placed in two cups, dissolved in 5.3 and transferred to volumetric flasks with a capacity of 100 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена. There is introduced shown in table 1 are the volumes of standard solutions (N 1, 3, and 2, 4), adjusted to the mark with water and mix. The thus obtained solutions A and B (regulirovanie solutions) is used to control the stability of the calibration parameters in the analysis of the steels with a mass fraction of tungsten and molybdenum in the range from 0.010 to 1.00%.

5.6.2 In two volumetric flasks with a capacity of 100 cm toГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаmake 10 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof the solution in the reference experiment and shown in table 1 are the volumes of standard solutions (N 1, 3, and 2, 4), adjusted with hydrochloric acid (1:9) to the mark and mix. So get regulirovanie solutions C and d, which are used to control the stability of the calibration parameters in the analysis of the steels with a mass fraction of tungsten and molybdenum in the range from 1.0 to 5.0%.

Regulirovanie the solutions were prepared with each batch of analyzed samples.

To prepare the entire series of solutions of analyzed samples of the bonds, the control of experience and recalibrating solutions using reagents from the same batch.

6 ANALYSIS

6.1 stability Control calibration parameters (recalibration) is performed before starting the measurement in accordance with the mathematical software tool. For control stability at the top of the range used regulirovanie solutions A and b, and at the bottom respectively regulirovanie solutions B and d in the analysis of the steels with a mass fraction of tungsten and molybdenum, are listed in table 1. For this purpose, conduct 3−5 measurements of the intensities of tungsten and molybdenum on the chosen analytical lines, spraying in plasma the corresponding regulirovanii solution.

Allow to use other methods of monitoring the stability of calibration parameters under mathematical software spectrometer setup.

6.2 Control of the correctness of the results of the analysis carried out by measurement of the mass fraction of tungsten and molybdenum in a solution of the bonds at 5.4.

In plasma spraying the appropriate solution, carry out three parallel measurements. The mass fraction that represents the average result of these measurements, is one of parallel measurements certified value of tungsten and molybdenum in GSO.

The average value of the two thus obtained parallel definitions mass fraction of tungsten and molybdenum should not vary from certified more than allowable value given in table 2. Otherwise, repeat the recalibration.


Table 2 — Norms of accuracy and norms of accuracy control of determination of mass fractions of tungsten and molybdenum

Percentage

         
Mass fraction of element

The error analysis results ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена

Permissible discrepancies
   

between the results of two parallel measurements ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена

between the results of the analysis carried out in different conditions, ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена

between the results reproduced
and certified values of the standard sample ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена

Tungsten
From 0.01 to 0.02 incl.
0,006
0,006
0,008
0,004
SV. 0,02 «0,05 «
0,008
0,008
0,010
0,005
«0,05» 0,10 «
0,013
0,014
0,018
0,009
«To 0.10» to 0.20 «
0,020
0,020
0,025
0,013
«To 0.20» to 0.50 «
0,030
0,030
0,040
0,020
«To 0.50» and 1.0 «
0,050
0,050
0,060
0,030
«1,0» 2,0 «
0,060
0,060
0,080
0,040
«To 2.0» 5,0 «
0,10
0,10
0,13
0,060
Molybdenum
From 0.01 to 0.02 incl.
0,004
0,004
0,005
0,0024
SV. 0,02 «0,05 «
0,006
0,006
0,007
0,004
«0,05» 0,10 «
0,009
0,009
0,011
0,007
«To 0.10» to 0.20 «
0,015
0,016
0,018
0,010
«To 0.20» to 0.50 «
0,024
0,025
0,030
0,016
«To 0.50» and 1.0 «
0,030
0,030
0,040
0,020
«1,0» 2,0 «
0,040
0,040
0,060
0,030
«To 2.0» 5,0 «
0,070
0,070
0,10
0,050

6.3 measurement of the mass fraction of tungsten and molybdenum is carried out for each solution analyzed samples similar to the procedure specified in 6.2. With this purpose, is sprayed into the plasma solutions the mass concentration of 1 mg/cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаin the analysis of the steels with a mass fraction of tungsten and molybdenum, up to 1.00% and solution concentration of 100 µg/cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаin the analysis of the steels with a mass fraction of tungsten and molybdenum from 1.00 to 5.0%.

6.4. After each measurement the system was washed by spraying hydrochloric acid solution (1:9).

7 PROCESSING OF RESULTS

7.1 Mass percent of tungsten and molybdenum in the sample ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаin % is calculated by the formula

ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена, (1)


where ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена — mass fraction of tungsten and molybdenum in the solution of the sample, %;

ГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибдена — weight of steel contained in 1 cmГОСТ Р 51056-97 Стали легированные и высоколегированные. Атомно-эмиссионный спектральный метод определения вольфрама и молибденаof solution, mg.

The final result of the analysis be the arithmetic mean of two parallel definitions for the two test portions of the test sample.

7.2 Standards of accuracy and standards for monitoring the accuracy of determining the mass fraction of tungsten and molybdenum are shown in table 2.

7.3 allowable Absolute discrepancies in the results of parallel measurements at p = 0.95 does not exceed the permissible differences for the respective concentrations shown in table 2. If you get results with differences more acceptable the analysis should be repeated using a new sample of the analyzed sample. If re-analysis of the differences between the results of parallel measurements once again exceeds the limit, the sample is rejected and replaced with a new.