GOST 17261-2008 GOST 3778-98 GOST 3640-94 GOST 25284.8-95 GOST 25284.7-95 GOST 25284.6-95 GOST 25284.5-95 GOST 25284.4-95 GOST 25284.3-95 GOST 25284.2-95 GOST 25284.1-95 GOST 25284.0-95 GOST 25140-93 GOST 23957.2-2003 GOST 23957.1-2003 GOST 23328-95 GOST 22861-93 GOST 21438-95 GOST 21437-95 GOST 19424-97 GOST 15483.10-2004 GOST 1293.0-2006 GOST 1219.1-74 GOST 1219.3-74 GOST 21877.6-76 GOST 21877.0-76 GOST 9519.1-77 GOST 15483.1-78 GOST 15483.0-78 GOST 1293.0-83 GOST 1293.3-83 GOST 26880.1-86 GOST 1219.4-74 GOST 1219.8-74 GOST 1219.2-74 GOST 860-75 GOST 21877.3-76 GOST 21877.1-76 GOST 21877.9-76 GOST 21877.4-76 GOST 21877.7-76 GOST 21877.2-76 GOST 21877.10-76 GOST 21877.8-76 GOST 22518.2-77 GOST 22518.4-77 GOST 9519.2-77 GOST 22518.1-77 GOST 1293.6-78 GOST 15483.11-78 GOST 15483.8-78 GOST 15483.3-78 GOST 15483.6-78 GOST 19251.3-79 GOST 20580.8-80 GOST 20580.2-80 GOST 20580.3-80 GOST 1293.11-83 GOST 1293.1-83 GOST 27225-87 GOST 30608-98 GOST 19251.7-93 GOST P 51014-97 GOST 17261-77 GOST 22518.3-77 GOST 9519.3-77 GOST 8857-77 GOST 15483.4-78 GOST 19251.0-79 GOST 19251.5-79 GOST 19251.2-79 GOST 20580.1-80 GOST 20580.6-80 GOST 20580.7-80 GOST 20580.4-80 GOST 1292-81 GOST 9519.0-82 GOST 1293.10-83 GOST 1293.12-83 GOST 1293.5-83 GOST 1293.2-83 GOST 30082-93 GOST 1219.6-74 GOST 1219.0-74 GOST 1219.5-74 GOST 1219.7-74 GOST 21877.5-76 GOST 21877.11-76 GOST 15483.9-78 GOST 15483.7-78 GOST 15483.2-78 GOST 1293.9-78 GOST 15483.5-78 GOST 19251.1-79 GOST 19251.6-79 GOST 19251.4-79 GOST 20580.0-80 GOST 20580.5-80 GOST 1293.7-83 GOST 1293.13-83 GOST 1293.14-83 GOST 1293.4-83 GOST 26880.2-86 GOST 26958-86 GOST 1020-97 GOST 30609-98 GOST 1293.15-90 GOST 1209-90 GOST 1293.16-93 GOST 13348-74 GOST 1320-74 GOST P 52371-2005

GOST 9519.0−82 calcium Babbits. General requirements for methods of spectral analysis (with Change No. 1)


GOST 9519.0−82

Group B59

STATE STANDARD OF THE USSR

CALCIUM BABBITS

General requirements for methods of spectral analysis

Lead-calcium bearing alloys. General requirements for methods of spectral analysis

AXTU 1709*
________________
* Introduced advanced Edit. N 1.

Date of introduction 1984−01−01


DEVELOPED by the Ministry of nonferrous metallurgy of the USSR

PERFORMERS

Kupryakov Y. P., A. A. Reznikov, L. I. Tunin, R. P. Petrov, V. I. Petrov, N. And. Abrosimov, A. E. Gruzdeva, T. A. Slobodian

INTRODUCED by the Ministry of nonferrous metallurgy of the USSR

Member of the Board of A. P. Snurnikov

APPROVED AND put INTO EFFECT by Decision of the USSR State Committee for standards, dated 3 September 1982, N 3518

The validity period is set with 01.01.84 to 01.01.89*
________________
* Expiration removed by Protocol No. 3−93 Interstate Council for standardization, Metrology and certification (I & C N 5−93). — Note the manufacturer’s database.


The Change N 1, approved and put into effect by the Decree of Gosstandart of the USSR from 25.06.1987 N 2467 c 01.03.1988

Change No. 1 made by the manufacturer of the database in the text ICS N 10, 1987

1. This standard establishes General requirements for methods of spectral analysis.

2. Analysis of the investigated samples is carried out using the three standards with the use of cast metal or of a synthetic standard samples of the enterprise and standard solutions.

3. The analyzed samples are taken by casting in a metal mold from molten metal or GOST 1209−78*.
________________
* On the territory of the Russian Federation the document is not valid. Standards 1209−90. — Note the manufacturer’s database.

4. Standard samples of the enterprise meet the requirements of GOST 8.315−78*. Composition of standard samples of the enterprise establish chemical methods in at least two laboratories according to GOST 1219.0−74-GOST 1219.8−74.
________________
* On the territory of the Russian Federation the document is not valid. Valid GOST 8.315−97. — Note the manufacturer’s database.

5. Analysis of standard samples of the enterprise for testing and graphing is carried out simultaneously with the analyzed samples.

6. Weighing is carried out with a maximum error of 0.0002 g.

7. Jobs and rules for their placement, utility rooms must meet the requirements of GOST 12.2.032−78, GOST 12.2.033−78.

8. Laboratory spaces must be equipped with ventilation system and to meet the requirements of GOST 12.4.021−75 with the provision of sanitary-hygienic requirements for working zone air according to GOST 12.1.005−76.

9. Electrical installations and electrical appliances used in the process of spectral analysis shall conform to the electrical codes and the requirements of GOST 12.2.007.0−75 and GOST 12.2.007.9−75.

Operation of electrical installations and electrical appliances must be in accordance with the requirements of GOST 12.1.019−79, rules of technical operation of electrical installations of consumers and safety of operation of electrical installations, approved by Gosenergonadzor USSR.
________________
* On the territory of the Russian Federation the document is not valid. Apply cross-industry Rules on labor protection (safety rules) for electrical installations" (SWEAT P M-016−2001), «Rules on labor protection at operation of electroinstallations». — Note the manufacturer’s database.

10. Spectrographs and spectrometers must be installed in separate rooms and located not less than 1.5 m from each other. Tripods spectrograph, flames and machines for sharpening carbon electrodes must be equipped with protective screen and local fume extraction.

11. Work tables and fume hood for working with chemicals should have the boards and shall be covered with fireproof materials and materials resistant to acids.

12. Formed in the excitation spectra of the electrical discharges, ozone, carbon monoxide, oxides of nitrogen and applied in the process, the antimony oxide, ethyl alcohol, the oxides of lead are toxic and have harmful effects on the human body. When working with these substances should be guided by the safety requirements according to GOST 12.1.007−76.

13. The content of toxic substances in the atmosphere of working zone shall not exceed: for ozone is 0.1 mg/mГОСТ 9519.0-82 Баббиты кальциевые. Общие требования к методам спектрального анализа (с Изменением N 1)carbon monoxide — 20 mg/mГОСТ 9519.0-82 Баббиты кальциевые. Общие требования к методам спектрального анализа (с Изменением N 1), NOx — 5.0 mg/mГОСТ 9519.0-82 Баббиты кальциевые. Общие требования к методам спектрального анализа (с Изменением N 1)in accordance with GOST 12.1.005−76.

Control of content of harmful substances in the air of working zone should be carried out in accordance with GOST 12.1.005−76 and GOST 12.1.007−76.

(Changed edition, Rev. N 1).

14. Acetylene cylinders should be placed outside the building in accordance with the rules and requirements for the installation and operation of gas pipelines, approved by Gosgortekhnadzor of the USSR.

Pipelines must comply with the requirements of safety rules in the gas sector* approved by Gosgortekhnadzor of the USSR, GOST 14202−69.
_______________
* On the territory of the Russian Federation the document is not valid. There are Federal rules and regulations in the field of industrial safety «safety Rules of gas distribution networks and gas consumption»; Federal norms and rules in the field of industrial safety «safety Rules for facilities using liquefied hydrocarbon gases»; Federal norms and rules in the field of industrial safety «safety Rules filling stations gas fuel». — Note the manufacturer’s database.

15. Fire safety requirements should correspond to GOST 12.1.004−85. The types of fire fighting equipment (fire extinguishers carbon dioxide type OU-2 according to GOST 7276−77, a box with sand and shovel according to GOST 3620−76), its placement and content must comply with GOST 12.4.009−83.

(Changed edition, Rev. N 1).

16. Collective means of protection must meet the requirements of GOST 12.4.011−75, GOST 12.3.002−75 and normative-technical documentation.

17. When working in the spectral laboratory should apply personal protective equipment: robes of cotton according to GOST 12.4.131−83, rubber gloves according to GOST 20010−74*, goggles-type (3N).
_______________
* On the territory of the Russian Federation the document is not valid. Standards 20010−93. — Note the manufacturer’s database.


For first aid in the laboratory should be a first aid kit with necessary medicines.

(Changed edition, Rev. N 1).

18. To work in the spectral laboratory allowed a person trained in safe work in accordance with GOST 12.0.004−79.